Introduction
Working Principle
Technical data
Vial size |
10ml, 13ml and 15ml vial |
Production capacity |
≥200 vials/min for 10ml vial |
Vial broken ratio |
≤0.01% |
Sterilization tunnel |
300℃—350℃ |
Sterilization time |
≥5min |
Temperature difference between heating and insulation stage |
≤±5-10℃ |
Air pressure balance accuracy |
±2Pa |
Vial temperature after sterilized |
<35℃ (or 5-10℃ higher than room temperature) |
Belt width |
800mm |
Belt speed |
100-400mm/min (frequency converter speed control) |
Ultrasonic frequency |
1000W |
Air exhaust volume in pre heating |
About 700m3/h |
Air inlet volume in pre heating |
About 300m3/h |
Air inlet volume in sterilization stage |
About 300m3/h |
Air exhaust volume of air pressure balancer |
About 300m3/h |
Fresh air supply |
1000-3000m3/h |
Chilled water supply |
7℃,1—3m3/h |
Power |
380V 50Hz;About 78KW; |
Outside dimension |
4290×2050×2450mm |
Noise |
≤70Db |
Weight |
About 4000Kg |